KLY-3/S SPINNER
KAPPABRIDGE
|
|
KLY-3/S Magnetic Susceptibility/Anisotiropy Systems The world's most sensitive commercially available laboratory instruments for measuring bulk magnetic susceptibility and anisotropy of magnetic susceptibility. Sensitivity: 20 x 10^-9 emu. The KLY-3 Static version measures the AMS of a static specimen on the same principle as the former KLY-2 instrument but is equipped with a fully automatic zeroing system (in both components) and has better sensitivity. The speci- men susceptibility is measured in on-line mode with the PC computer in 15 different positions following rotatable design. The positions are changed manually and, using special software, the susceptibility tensor is calculated including the statistical errors in its determination. The KLY-3S Spinner version measures the AMS
of a slowly spinning specimen rather than a static specimen and
one has to adjust the specimen only in three perpendicular positions.
The measurement is rapid (about 2 minutes per specimen) and precise,
profiting from many susceptibility determinations in each plane
perpendicular to the axis of specimen rotation. New software combines
the measurements in three perpendicular planes plus one bulk value
to create a complete susceptibility tensor. The errors in determination
of this tensor are estimated using new method based on multivariate
statistics principle. |
|
SPECIFICATIONS: Specimen Size: ..................................
25.4 mm dia. x 22 mm Cylindrical Operating Frequency:...................... 870 Hz Field Intensity: .................................. 300 A/m Field Homogeneity:........................... 0.2% Measuring Ranges: ........................... 0.1, 0.01, 0.001, 0.0001 (SI) Sensitivity (typical): .........................Bulk
30 x 10^-9 (SI), Aniso 30 x 10^-9 |
Call GISCO / 1-800-523-1526 (952-929-8000)
FAX 952-926-5498 / EMAIL:sales@giscogeo.com
Back to paleomagnetic index
![]()